Vicat Softening Point Temperature Testing Equipment Microcontroller for Polymer Materials

Brand Name:DAXIAN
Certification:CE , SGS
Model Number:DX8534
Minimum Order Quantity:1 set
Delivery Time:30 days after order confirmed
Payment Terms:L/C,D/A,D/P,T/T,Western Union,MoneyGram,In cash
Contact Now

Add to Cart

Verified Supplier
Location: Dongguan Guangdong China
Address: No.168 Zhen'an West Road Changlian Industry Mansion Xiabian Community Dongguan China
Supplier`s last login times: within 28 hours
Product Details Company Profile
Product Details

Microcontroller Thermal Deformation, Vicat Softening Point Temperature Testing Equipment for Polymer Materials


Scope of application:

Mainly used for thermal deformation of polymer materials (such as plastics, rubber, nylon, electrical insulation materials, etc.) and determination of Vicat softening point temperature. Applicable to universities, research institutes and product quality supervision and inspection units.


Compliance with the standard:

in line with GB/T 8802-2001, ISO 2570 hard polyethylene (PVC), GB/T 1633, GB/T 1634, ISO 75-1987 (E), ISO 306-1987 (E) and other standards.

Standards compliant:

GB/T 1633-2000 (idt ISO 306:1994) Determination of Vicat Softening Temperature (VST)

GB/T 1634.1-2004 (idt ISO75-2:2003) Determination of plastic deformation temperature (general test method)

GB/T 1634.2-2004 Plastics Determination of Deflection Temperature at Loading (Plastics, Hard Rubber and Long Fiber Reinforced Composites)

GB/T 1634.3-2004 Plastics Determination of Deflection Temperature at Loading (High Strength Thermosetting Laminate)

GB/T 8802-2001 (eqv ISO 2507:1995) Determination of Vicat softening temperature of thermoplastic pipes and fittings


Parameters:

Test range: room temperature ~ 300 °C;

Number of sample holders: 2, 3, 6;

Heater power: 4KW;

Vicat variability: 1mm.

Deformation resolution error: 0.005mm

Heating rate: A = 5 °C / 6min B = 12 °C / 6min;

Maximum temperature measurement error: ±0.5°C

Maximum temperature control error: ±12°C/6 minutes (thermal deformation test) ±5°C/6 minutes (Vicat test)

1# weight (ie load rod, standard indenter, sensor probe, tray, screw): 68g ±1g

Heating medium: Silicone oil (200 cSt or less, flash point above 300°C)

Cooling method: natural cooling or water cooling (C type machine)

Dimensions: 520×420×1100mm;

Main power supply: 220V+10%, 15A, 50HZ;