Add to Cart
KE8302M-1000 is used for testing and calibration of single / three phase clamp CT meter, including on-site installed meter and portable standard meter.
| KE8302-1000-3 | KE8302-1000-6 | |
| Number of meter position | 3 | 6 |
| Accuracy | ± 0.05% | |
| Voltage output | ||
| Output range (P-N) | 3×(24V - 300V) | |
| Output power of voltage | 3×100VA | 3×200VA |
| Resolution | Better than ± 0.01% | |
| Stability | Better than ± 0.02% / 150s | |
| Load regulation rate | Better than ± 0.01% | |
| Distortion factor | Better than ± 0.3% (linear resistance load) | |
| Harmonics output | 2nd -21st free programmable | |
| Current output | ||
| Output range | 3×(0.5A - 1000A) | |
| Output power of current | 3×400VA | 3×800VA |
| Resolution | Better than ± 0.01% | |
| Stability | Better than ± 0.02% / 150s | |
| Load regulation rate | Better than ± 0.01% | |
| Distortion factor | Better than ± 0.3% (linear resistance load) | |
| Harmonics output | 2nd -21st free programmable | |
| Phase output | ||
| Output range | 0 -- 360° | |
| Resolution | ± 0.01° | |
| Frequency output | ||
| Output range | 45Hz - 65Hz | |
| Resolution | ± 0.01Hz | |
| Operation conditions | ||
| Power supply | 3×220/380V±10% ,50/60Hz±2Hz | |
| Temperature | -10°C -- +40°C | |
| Humidity | 35% -- 85%RH | |