FDA Multi-Drug Test Dipcard in Urine Drug of Abuse Rapid One Step Test Kit
【INTENDED USE】
2-15 combination of the following drugs:AMP/ BAR/ BUP/ BZO/ COC/
COT/ ETG/ FYL/ K2/ KET/ MAM/ MDMA/MET/ MOP300/ MQL Rapid Multi-Drug
2-15 Test is a rapid, screening test for the qualitative detection
of multiple drugs and drug metabolites in human urine at specified
cut off levels. For professional use only. For in vitro diagnostic
use only.
DOA (Identifier) | Calibrator | Cut-off lever |
Cocaine (COC) | Benzoylecgonine | 300 ng/mL |
Amphetamine (AMP) | d-Amphetamine | 1000ng/mL |
Marijuana (THC) | 11 9-THC-9-COOH | 50 ng/mL |
Benzodiazepines (BZO | Oxazepam | 300 ng/mL |
Opiate 2000 (OPI) | Morphine | 2000ng/mL |
【OPERATION】
Test must be in room temperature (15ºC to 30ºC)
1. Donor collects urine specimen in a urine cup.
2. Open the sealed pouch by tearing along the notch. Remove the
test cassette from the pouch and place it on a level surface.
3. Hold the sample dropper vertically, and add exactly three drops
of the urine specimen into the sample well.
4. The result should be read at 5 minutes. Do not interpret the
result after 10 minutes. See the illustration below.
【INTERPRETATION OF RESULTS】
Preliminary positive (+) Only one colored band appears, in the
control region (C). No colored band appears in the test region (T).
Negative (-) Two colored bands appear on the membrane. One band
appears in the control region (C) and another band appears in the
test region (T).
Invalid Control band fails to appear. Results from any test which
has not produced a control band at the specified read time must be
discarded.
Please review the procedure and repeat with a new test. If the
problem persists, discontinue using the kit immediately and contact
your local distributor.
【CONTENTS】
Drug Test
Desiccant
Leaflet with instruction for use
【STORAGE AND STABILITY】
The kit should be stored at 2-30°C until the expiry date printed on
the sealed pouch
The test must remain in the sealed pouch until use
Keep away from direct sunlight, moisture and heat
Do not freeze.
Care should be taken to protect the components of the kit from
contamination.
Do not use if there is evidence of microbial contamination or
precipitation.
Biological contamination of dispensing equipment, containers or
reagents can lead to false results.