Hematology Analyzer Reagent SWELAB AC970 CBC 3 Part Diluent Rinse
Lyse Cleaner
Specification of Swelab Reagents
Compatible Reagents on Swelab AC970 AC920 AC910 AC900 |
Cat. No | Description | Pack Size |
DWX-31301 | DILUEN- AC | 20L |
DWX-31302 | LYSE-AC | 5L |
DWX-31303 | RINSE-AC | 5L |
Intended Use of Hematology Reagents To be used for blood cell counting, sizing, and WBC
differentiation, hemoglobin determination on SWELAB AC970
Hematology Analyzers.
Close system: The manufacturers impose restrictions on software by scanning
their barcode when changing reagents to prevent other compatible
reagents.
Open system: The manufacturers do not impose restrictions on software so that
reagents can be used no matter what brands.
Storage of Swelab CBC Reagents
Store the product at 2-35℃ in a dust free place with good
ventilation. Temperature outside the range may damage the product.
The open vial life of the product is 60 days. Unused reagents
should be discarded after 60 days. Do not mix residual with a new
carton.
Should the Blood Cell Counter instrument do a calibration before
using our reagents?
First of all,if your instrument used the original reagents before
using our products, you needn’t to do calibrate.
If your instrument used other compatible reagents, calibration is
necessary before using our reagents.
Limitations of Detection Method
If the temperature of the reagent is out of the specified range,
the specimen's measurement parameters may appear abnormal. If in
this condition, please confirm the measurement parameters with the
microscopy method. Please refer to the instrument operation manual
for details.
Storage
The product should be stored in 2℃-35℃ and the storage period is 2
years. After opening the bottle, it is used in 15℃-30℃, and the
validity period is 60 days.
This product is for In Vitro Diagnostic Use only.
1) What are the reagents of hematology Analyser?
At present, hematology analyzers on the market can be divided into
3 part hematology analyzers and 5 part hematology analyzers.
Hematology reagents are also indispensable in the use of hematology
analyzer. Hematology reagents include hemolytic agents, diluents,
cleaning solutions and concentrated cleaning solutions.
2) What is the diluent in the hematology analyzer?
The diluent for the hematology analyzer comprises sodium chloride,
sodium sulfate, a phosphate buffer agent, ethylene diamine
tetraacetic acid, 1-pyridone-2-sulphur and formaldehyde, and the pH
value of the solution is 6.5-7.4, and preferably is 7.2; the
content of each component is preferably as follows: 5g/L of ...
3) Which diluent is good for RBC and platelet counting?
For this, the blood specimen is diluted (usually in 1:200 ratio)
with the help of RBC diluting fluid (commonly the Hayem's Fluid)
which preserve and fix the Red blood cells. The Hayem's fluid is
isotonic to the Red blood cells and does not cause any damage to
it.
4) What is the use of Lyse in hematology analyzer?
It is used to lyse (destroy) red blood celis for hemogiobin
determinations or aid in the counting of white blood cells.
Applicable to the hematology analyzers.
5) What is the composition of Lyse for hematology analyzer?
The lytic reagent composition comprises a short chain alkyl
oxyethanol selected from the group consisting of 2-methoxyethanol,
2-ethoxyethanol, 2-propoxyethanol, and 2-isopropoxyethanol in a
sufficient amount to preserve leukocytes; a non-lysing nonionic
surfactant as a debris solublizer; and an inorganic buffer to ...