IEC61010-1 Clause 10.2.2 Electrical Testing Equipment Cord Anchorage Push-pull and Torque Tester

1. Introduction

The electrical testing equipment, named Power Cord Push-Pull and Torque tester, is designed and manufactured according to the requirements of IEC62368-1 Appendix G.7.3.2.1 and Table G.8, IEC60335-1 clause 25.15, IEC60974-1 clause 10.5, GB14711 clause 25.8, IEC60884-1 clause 23.1 and Table 18, IEC60598-1 clause 5.2 and Table 5.2, IEC60601-1 clause 8.11.3.5 and Table 18, IEC61010-1 clause 6.10.2.2. It is also widely used in power cord tension and torque tests on lamps, home appliances, IT/AV appliances, and electrical accessories. The equipment is equipped with a color touch screen, pull-push weights (according to standard configuration), and torque weights (according to standard configuration).


2. Standard

IEC62368-1 Audio/video, information and communication technology equipment Part 1: Safety requirements Annex G.7.3.2.1 and table G.8

IEC60335-1 Household and similar electrical appliances - Safety - Part 1: General requirements Clause 25.15

IEC60974-1, GB15579.1 Arc welding equipment - Part 1: Welding power sources Clause 10.5

GB14711 Clause 25.8

IEC60884-1 Plugs and socket-outlets for household and similar purposes -Part1: General requirements Clause 23.1 and 18

IEC60598-1 Luminaires – Part 1: General requirements and tests Clause 5.2.10.3 and table 5.2

IEC60601-1 Medical electrical equipment-Part 1: General requirements for basic safety and essential performance Clause 8.11.3.5 and table 18

IEC61010-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance Clause 6.10.2.2 and table 11


3. Parameters

Working PowerAC 220V/50Hz, 100W max.
Control system5-inch color touch screen
Pull force weight20N*2, 30N*2 (can realize 30N, 60N, 80N, and 100N pull forces)
Torque weight5N*1, 125.N*1, 17.5N*1(can realize 0.1Nm, 0.25Nm and 0.35Nm torque)
Test times0~99999 presetable, default 25 times
Test speedAdjustable
Torque time0~99H99M99S
FixtureMultiple screw holes for locking, screw compression locking, platform can be raised and lowered
StructureAluminum bracket, aluminum-plastic composite structure

4. Pull force and torque Table

Mass of appliance (kg)Pull force (N)Torque (N.m)
≤1300.10
>1 ≤4600.25
>41000.35

5. Figure

China IEC61010-1 Clause 10.2.2 Electrical Testing Equipment Cord Anchorage Push-pull and Torque Tester supplier

IEC61010-1 Clause 10.2.2 Electrical Testing Equipment Cord Anchorage Push-pull and Torque Tester