Product Details

Spring Hammer Calibration Device Calibrator as per IEC60068-2-75 Annex B.3 and figure B.5


1. Introduction

The spring hammer calibration device calibrator is designed according to IEC60068-2-75 Annex B.3 and figure B.5 for checking the impact energy accuracy of spring hammer calibration device. The calibation is made with an impact energy of 1J, which is achieved with a height of fall of 408±1mm, with a calibation striking element of 250g.


2. Parameter

Accuracy0.01J
Repeated Accuracy±0.01J
Max measuring range0-2J
Calibration striking element250g
Reference pointAn impact energy of 1J is achieved with a height of 408±1mm
Pendulum rope center distance2m