Surface Scattering Light Turbidity Detector with 0-2000 NTU Range and LED Light Source for Water Quality Analysis

Brand Name:Yokogawa
Model Number:TB830D
Minimum Order Quantity:1pcs
Delivery Time:5-8 work days
Payment Terms:L/C,D/A,D/P,T/T,Western Union,MoneyGram
Supply Ability:1000 Piece/Pieces per Week negotiable
Contact Now

Add to Cart

Verified Supplier
Address: i City, No11, TangYan South road, Yanta District, Xi'an,Shaanxi,China.
Supplier`s last login times: within 16 hours
Product Details Company Profile
Product Details

"Reduce maintenance"
 "Digital solution for water quality"

Many of our customers face these challenges - Yokogawa Turbidity Analyzers TB800 series resolve them.

Yokogawa's turbidity meters have been developed and improved to suit various applications in water treatment plant turbidity measurement. As a result, we have accumulated many achievements and gained the trust of our customers since our first product was launched in 1959.

Japan’s most trusted turbidimeter*.

The TB830D is part of the SENCOM4.0 platform family and can connect to FLXA402T. Thanks to a smart unit equipped detector, TB830D gives you more insight into the ordinary operation.

The TB830D is a turbidity analyzer that equipped LEDs as the light source, is dirt-resistant, and offers reliable measurements and reduced OPEX due to the addition of advanced diagnostic functions.

*Based on a Yokogawa survey in 2019


Less susceptible to contamination due to non-contact with the sample

Compatible with various applications

  • Surface scattering measurement method minimizes contamination
  • LED light source for 3-year long life and reduced maintenance costs
  • High sensitivity and high accuracy with long-standing reliability
  • Good linearity and reproducibility
  • Measuring range from 0-2 NTU to 0-2000 NTU
  • Improved cleaning system reduces maintenance
  • Easy-to-clean cell structure (by GS)

Multi sensor input
Not only Turbidity Detector but also Chlorine Sensor Unit, pH sensor and conductivity sensor are connectable to one transmitter.

Application

Turbidity of tap water, sewage, river water and water used in general processes.

Water Purification Treatment

  • Sedimentation
  • Flocculation
  • Settling Basins
  • Clean water reservoir
  • Distributing pump
  • Distribution tank & Piping

Wastewater Treatment

  • Discharge/Reuse

Specifications

Measuring methodSurface scattering method
Measuring Range
(Turbidity standard: Formazin)
0-2 to 0-2000 NTU
Sample ConditionsTemperature0 to 50°C
Flow Rate1.5 to 2 L/min
Pressure500 kPa max
Power Supply100 to 240 V DC, 50/60 Hz
Digital CommunicationModbus TCP/IP, Modbus RTU
InstallationIndoor (Outdoor use requires protection against rain.Direct sunlight must be avoided.)
Cleaning systemAutomatic wash


Performance

Linearity0-1000 NTU: ±2% of reading or ±0.04 NTU whichever is greater
over 1000 NTU: ±4% of reading
Repeatability±1% of reading or ±0.02 NTU whichever is greater