Operating Environment-Friendly KGD WQ 501 Engine Emission Testing System for Accurate Results

Brand Name:KACISE
Certification:CE
Model Number:KGD-WQ-501
Minimum Order Quantity:1PCS
Delivery Time:5-8 work days
Payment Terms:L/C,D/A,D/P,T/T,Western Union,MoneyGram
Contact Now

Add to Cart

Verified Supplier
Address: i City, No11, TangYan South road, Yanta District, Xi'an,Shaanxi,China.
Supplier`s last login times: within 16 hours
Product Details Company Profile
Product Details

KGD - WQ - 501 Engine Emission Testing System for Engines

Product Introduction

The KGD - WQ - 501 Engine Emission Testing System is an integrated system device specifically designed for measuring conventional gas components such as THC, CH4, NOx, CO2, CO, and O2 in full - flow dilution or raw exhaust emission tests in laboratories. It includes functional modules such as a sampling unit, a pre - treatment unit, and various gas analyzers, and is applicable to research, development, and certification tests of all types of engines and fuels.

Product Features

  • Adopts self - developed patented gas sensing technologies such as NDIR, NDUV, CLD, and HFID, with one - stop service throughout the life cycle.
  • The gas analyzer has a wide measurement range, suitable for all laboratory emission test application scenarios including ultra - low emission levels.
  • The gas analyzer has high accuracy and fast response, enabling high - precision continuous online measurement of diluted exhaust and raw exhaust.
  • Can be equipped with a multi - channel multifunctional sampling unit to flexibly respond to multi - point sampling measurement of exhaust pipes and EGR rate measurement.

Technical Parameters

Gas ComponentTHCCH4NOxCOCO2O2
Measurement PrincipleHFIDNMC - FIDNDUV/CLDNDIRNDIRMPD
Minimum Range0 - 10 ppmC0 - 10 ppmC0 - 10 ppm0 - 50 ppm0 - 0.5%0 - 1%
Maximum Range0 - 30000 ppmC0 - 3000 ppmC0 - 10000 ppm0 - 10%0 - 20%0 - 25%
Response TimeT10 - 90 < 2.5s
Accuracy≤ ±2.0%WR or ≤ ±0.3%FS
LinearityDetermination coefficient: ≥ 0.998; Standard error: ≤ 1%FS; Slope: 0.99 - 1.01; Intercept coefficient: ≤ 0.5%FS
Repeatability≤ ±0.5%FS
Drift≤ ±1%FS/8h
Operating EnvironmentAmbient temperature: 5 - 40°C; Ambient humidity: ≤ 80%RH