KWS-700 Online pH Sensor
Introduction
KWS-700 online pH sensor is used in environmental water quality
monitoring, acid/alkali/salt solution, chemical reaction process,
industrial production process, it can meet the requirments of
online pH measurement for most industrial applications.
Feature
Signal output: RS-485 (Modbus/RTU protocol).
Convenient connection to third-party equipment such as PLC, DCS,
industrial control computer,general controller, paperless recording
instrument or tounch screen.
Double high-impedance differential amplifier with strong
anti-interference and fast response.
The patented pH probe, the internal reference solution oozes very
slowly from the microporous salt bridge under the pressure of at
least 100KPa (1Bar), and its forward bleed continues for more than
20 months. Such a reference system is very stable and the electrode
life is extended by a factor of two compared to conventional
industrial electrodes.
Easy to install: 3/4 NPT pipe thread for easy submersible
installation or installation in pipes and tanks.
Technical Specifications
Model | KWS-700 |
Measuring Range | 0~14pH |
Accuracy | ±0.1pH |
Resolution | 0.01pH |
Temperature Range | 0 ~ 50℃ |
Work Pressure | <0.2MPa |
Temperature Compensation | automatic temperature compensation |
Output | RS-485, MODBUS protocol |
Power Supply | 6~12VDC ±10% |
Wetted Part Material | POM |
Installation | Immersion mounting |
Cable Length | 5 m, customizable |
Calibration Method | Two-point calibration |
Ip Grade | IP68 |
Dimension
Installation and Electrical Connection
Installation
Note: The sensor should not be installed upside down or horizontally when installed, at least at an angle of 15 degrees or more.
Wiring
a) Red line - power cord (12 ~ 24V)
b) Black line - ground (GND)
c) Blue line - 485A
d) White line - 485B
e) Bare wire - shielded wire
After wiring is completed, it should be carefully checked to avoid
incorrect connections before powering up.
Cable specification: Considering that the cable is immersed in water (including sea water) for a long time or exposed to the air,the cable has certain corrosion resistance.The outer diameter of the cable is Φ6 mm and all interfaces are waterproof.
Maintenance
Use and maintenance
When measuring the pH sensor, it should be cleaned in distilled
water (or deionized water), and the filter paper should be used to
absorb moisture to prevent impurities from being introduced into
the liquid to be tested. 1/3 of the sensor should be inserted into
the solution to be tested.
The sensor should be washed when not in use, inserted into a
protective sleeve with a 3.5 mol/L potassium chloride solution, or
the sensor inserted into a container with a 3.5 mol/L potassium
chloride solution.
Check if the terminal is dry. If it is stained, wipe it with
absolute alcohol and dry it. Avoid long-term immersion in distilled
water or protein solution and prevent contact with silicone grease.
With a longer sensor, its glass film may become translucent or with
deposits, which can be washed with dilute hydrochloric acid and
rinsed with water. The sensor is used for a long time. When a
measurement error occurs, it must be calibrated with the meter for
calibration.
When the calibration and measurement cannot be performed while the
sensor is being maintained and maintained in the above manner, the
sensor has failed. Please replace the sensor.
Standard Buffer Ph Reference Table
The actual reading and standard of the instrument sometimes have an error of ±1 digit.
Calibration
Note: the sensor has been calibrated at the factory, if not exceeding the measurement error, the calibration should not be random.
a) Zero calibration
Use 250 mL of distilled water in a measuring cylinder,pour into a
beaker, add a packet of calibration powder with pH=6.86 , stir
evenly with a glass rod until the powder is completely dissolved,
configure the solution with pH=6.86 , put the sensor into the
solution, wait for 3 ~ 5 minutes, after the value is stable, see if
the displayed value is 6.86 . If not, you need to perform zero
calibration. Refer to the appendix for calibration instructions .
b) Slope calibration
For acidic solution: Take 250 mL of distilled water in a measuring
cylinder , pour into a beaker, add a packet of calibration powder
with pH=4.00 , stir evenly with a glass rod until the powder is
completely dissolved, and configure the solution topH=4.00 ; In the
solution, wait for 3 to 5 minutes. After the value is stable, see
if the value is 4.00 . If not, the slope calibration is required.
Refer to the appendix for the calibration instructions.
For alkaline solution: Take 250 mL of distilled water in a
measuring cylinder , pour into a beaker, add a packet of
calibration powder with pH=9.18 , stir evenly with a glass rod
until the powder is completely dissolved, and configure the
solution to pH=9.18 ; Into the solution, wait for 3 to 5 minutes.
After the value is stable, see if the display is 9.18 . If not, the
slope calibration is required. Refer to the appendix for the
calibration instructions.
Quality assurance
Quality inspection departments have standardized inspection procedures, with advanced detection equipment and instruments, and in strict accordance with the test procedures for products that do 72 hours aging test, stability test, not asubstandard products factory.
The receiving party directly returns the product batch with a failure rate of 2% , and all the costs incurred are borne by the supplier. The reference standard refers to the product description provided by the supplier.
Guarantee the quantity of goods and the speed of shipment.
Service commitment
The company provides local after-sales service within one year from
the date of sale, but does not include damage caused by improper
use. If repair or adjustment is required, please return it, but the
shipping cost must be conceited. Damaged on the way, the company
will repair the damage of the instrument for free.
Appendix data communication
1. Data Format
The default data format for Modbus communication is: 9600, n, 8, 1
(baud rate 9600bps, 1 start bit, 8 data bits, no parity, 1 stop
bit).
Parameters such as baud rate can be customized.
2. Information Frame Format
a) Read data instruction frame
06 03 xx xx xx xx xx xx
Address Function code Register address Number of registers CRC
check code (low byte first)
b) Read data response frame
06 03 xx xx...... xx xx xx
Address Function code Bytes Answer data CRC check code (low byte
first)
c) Write data instruction frame
06 06 xx xx xx xx xx xx
Address Function code Register address Write data CRC check code
(low byte first)
d) Write data response frame (same data command frame)
06 06 xx xx xx xx xx xx
Address Function code Register address Write data CRC check code
(low byte first)
3. Register Address
Note:
a) Modbus register address according to the protocol defined with
the register type register start address (16 hex represents the
actual starting address register in parentheses).
b) When changing the sensor address, the sensor address in the
return command is the changed new address.
c) The data definition that returns the measured value when reading
data:
xx xx xx xx xx xx Xx xx
2- byte measurement 2 bytes measurement value scale number 2 bytes temperature value 2 bytes temperature scale
The data type defaults to: double-byte integer, high byte first;
others are optional as floating-point types.
4. Command example
a) Read data instruction:
Function: Obtain the pH and temperature of the measuring probe; the
unit of pH is pH; the unit of temperature is °C.
Request frame: 06 03 00 00 00 04 45 BE ;
Response frame: 06 03 08 00 62 00 02 01 01 00 01 24 59
Example of reading:
pH: 00 62 indicates the hexadecimal reading pH value, 00 02
indicates that the pH value has 2
decimal places and is converted to a decimal value of 0.98 .
Temperature value: 01 01 indicates the hexadecimal reading
temperature value, 00 01 indicates that
the temperature value has 1 decimal place and is converted to a
decimal value of 25.7 .
b) Calibration instructions:
Zero calibration
Function: Set the pH zero calibration value of the electrode. The
zero value is based on the 6.86 pH standard. The examples are as
follows
Request frame: 06 06 10 00 00 00 8C BD
Response frame: 06 06 10 00 00 00 8C BD
Slope calibration
Function: Set the pH slope calibration value of the electrode; the
slope calibration is divided into high point and low point
calibration, and the alkaline solution is measured at the high
point; the acidic solution is measured at the low point, where the
standard solution is high here.
point 9.18pH, low 4.00pH standard solution of the calibration
reference, examples are as follows:
High point standard solution 9.18 pH calibration:
Request frame: 06 06 10 04 00 00 CD 7C
Response frame: 06 06 10 04 00 00 CD 7C
Low standard solution 4.00 pH calibration:
Request frame: 06 06 10 02 00 00 2D 7D
Response frame: 06 06 10 02 00 00 2D 7D
c) Set the device ID address:
Role: set the MODBUS device address of the electrode;
Change the device address 06 to 01. The example is as follows
Request frame: 06 06 20 02 00 01 E3 BD
Response frame: 06 06 20 02 00 01 E3 BD
5. Error response
If the sensor does not execute the host command correctly, it will return the following format information:
a) CODE : 01 – function code error
03 – data is wrong
b) COM : Received function code
Xi'an Kacise is certified by ISO 9001-2008, GJB 9001B-2009, CE,
intrinsic safety explosion-proof and isolated safety
explosion-proof, etc. which assure our products are conform to the
international standards.
We have been specializing in manufacturing pressure, level and flow
measuring instruments and related products for 10 years. We have
experienced technical and management personnel, through the efforts
of our entire staff, we have become a professional measuring
instrument manufacturer and solution provider. And we are
continuously trying to improve product quality.
Currently, Kacise sensor is already exported to America, Spain,
Australia, Chile and 20 other countries and regions, and received
many favorable feedbacks. Our overseas distributers are also
increasing. We regards “reasonable prices, efficient production
time and good after-sales service” as our tenet, and hope to
cooperate with more customers for mutual development.