Product Introduction
This PLA-800 Orthopedic plasma surgery system are designed
especially for Spine and Sports Medicine treatment. Such as knee,
shoulder, elbow, wrist, spine, cervical, ect.
Product Diagram
1 Plasma Generator | 2 Power Cord | 3 Footswitch Plate | 4 ABLATION Control |
5 PLACOAG Control | 6 Plasma Probe | 7 Receptacle of Footswitch | 8 Receptacle of Handle |
Working Principle
A 50 μm plasma area is formed around the electrode through a
conductive medium (mainly H2O, NaCl and air). Unlike traditional
electrosurgical methods, which separate and solidify tissues by
heating the tissue, Metron plasma knives use the kinetic energy
generated by the rapid movement of plasma in an electric field at
100 kHz to decompose the tissue, thereby achieving the purpose of
cutting and clotting.
Technical Parameter
Input Volts | 198-240V 50Hz 100-120V 60Hz Export device could convert automatically per customer requirements |
Output Power | Ablation or resection: ≤351W Coagulation or hemostasis: ≤111W |
Working Frequency | 100 KHz ± 10% |
Dimension | 405*440*130mm (L*W*H) |
Working Mode | Ablation and resection: level 1-10 Coagulation: level 1-10 |
Weight | 7.0kg |
Product Advanatages
Description | Remark |
Manufacturer Logo | MECHEN Brand. |
Pads | Insulation and support on the surface. |
Receptacle of Footswitch | Connects with Footswitch which determines ablation or coagulation mode. |
Receptacle of Treatment Handle | Connects with the treatment accessories. |
Power Switch | System start power button. “O” means power off; “I” means power on. |
ABLATION mode Selector | Select ABLATION mode 1 or 2 Output waveform. |
Indicator of Treatment Probe Connection | LED turns on green when the probe cable is connected and turns on red when it is not connected. |
Indicator of Footswitch Connection | LED turns on green when the footswitch cable is connected and turns on Red when it is not connected. |
ABLATION Power Setting Display | Shows the preset ABLATION output value in two digits. “01” with Min. power output, and “10” with the highest output. |
Ablation Output Power Up/Down Button | The number displayed indicates the ablation nominal power, in watts, which will be delivered to the patient when the mode is activated. Only available for changing in ready mode. |
System Features
1. Push-button operation, easy to use;
2. Real-time monitoring of the tissue impedance value during the
surgery to tell the degree of surgery;
3. peeling cutting probe(electrode), cutting is more convenient and
rapid;
4. more surgical electrodes to choose to meet more parts of tissue
ablation, cutting, coagulation and hemostasis;
5. ablation cutting mode 1, ablation cutting mode 2, to meet the
ablation cutting needs of different organizations.
Storage Condition
This device should be stored in a well-ventilated room without
corrosive gas. The height above ground and the distance from wall
should be above 30cm. Stack limit is 6 cases. Before being stored
for a long time, it should be cleaned thoroughly for proper packing
and storage. And it should also be taken out for power up every
year, in order to avoid it from moisture, mildew or even damage.
The suitable weather and environment condition is as below:
1. Temperature: -40℃~+55℃;
2. Relative humidity: ≤90%;
3. Atmospheric pressure: 50kPa~106kPa