multiple cassette 6 panel Drug Testing kits Saliva, Saliva Specimen, High Accuracy, easily and quick,


Intended Use


The Oral Fluid Drug Screen Device for AMP/MET/COC/OPI/THC/BZO is a lateral flow chromatographic immunoassay for the qualitative detection of amphetamine, methamphetamine, cocaine, opiates, THC, BZO and their metabolites in oral fluids at the following cut-off concentrations:


ProductCat NoSpecimenSensitivity
Drugs Testing Kits Saliva
AMPYD101S3Saliva1000 ng/mL
YD101C4
BARYD201S3Saliva300 ng/mL
YD201C4
BZOYD301S3Saliva300 ng/mL
YD301C4
COCYD401S3Saliva300 ng/mL
YD401C4
MOR(MOP)YD501S3Saliva300 ng/mL
YD501C4
METYD601SMSaliva500 ng/mL
YD601CM
MTDYD701S3Saliva300 ng/mL
YD701C4
MDMAYD801S3Saliva500 ng/mL
YD801C4
THCYD901S3Saliva50 ng/mL
YD901C4
BUPYD111S3Saliva100 ng/mL
YD111C4
PCPYD121S3Saliva25 ng/mL
YD121C4
K2YD131S3Saliva50 ng/mL
YD131C4
OPIYD141S3Saliva2000 ng/mL
YD141C4

This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) and gas chromatography/tandem mass spectrometry (GC/MS/MS) are the preferred confirmatory methods. Professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.


DIRECTIONS FOR USE


Allow the test, urine specimen, and/or controls to reach room temperature

(15-30ºC) prior to testing.

  1. Specimen Collection
  2. Using the provided oral fluid swab, sweep the inside of the mouth for 3 minutes. The sponge will gradually soften as oral fluid is absorbed, and should be completely soft after 3 minutes.
  3. Stop sweeping the mouth, and push the moistened swab firmly into the specimen dispenser. Push the collector into the collection chamber and press down firmly to release as much liquid as possible. Be sure that at least 0.5 mL liquid has been collected in order to make re-confirmation test possible, if necessary.
  4. Tightly shut the outer cap of the specimen dispenser.
  5. Test Procedure
  6. Remove the test from its sealed pouch, and place it on a clean, level surface. Label the test with patient or control identification. For best results, the assay should be performed within one hour.
  7. Unscrew the dispensing tip (not the outer cap) of the specimen dispenser, and transfer 3 drops of fluid (approximately 100 µL) to each specimen well (S) of the device and start the timer.
  8. Avoid trapping air bubbles in the specimen wells (S), and do not add any solution to the result areas.
  9. As the test begins to work, color will migrate across the membrane. 10. Wait for the colored band(s) to appear. The result should be read at 10 minutes. Do not interpret the result after 20 minutes.

INTERPRETATION OF RESULTS


NEGATIVE:* Two lines appear. One colored line should be in the control region (C), and another apparent colored line adjacent should be in the test region (Drug/T). This negative result indicates that the drug concentration is below the detectable level. *NOTE: The shade of color in the test line region (Drug/T) will vary, but it should be considered negative whenever there is even a faint line.


POSITIVE: One colored line appears in the control region (C). No line appears in the test region (Drug/T). This positive result indicates that the drug concentration is above the detectable level. INVALID: Control line fails to appear. Insufficient specimen volume or incorrect procedural techniques are the most likely reasons for control line failure. Review the procedure and repeat the test using a new test panel. If the problem persists, discontinue using the lot immediately and contact the manufacturer.



QUALITY CONTROL


A procedural control is included in the test. A line appearing in the control region

(C) is considered an internal procedural control. It confirms sufficient specimen volume, adequate membrane wicking and correct procedural technique.

Control standards are not supplied with this kit. However, it is recommended that positive and negative controls be tested as good laboratory practice to confirm the test procedure and to verify proper test performance.


China Multiple Cassette Mouth Swab Drug Test Kits Saliva Specimen For Various Drug supplier