Manual Pensky-Martens Closed-Cup Flash Point Apparatus
Summary
Cheap Manual Pensky-Martens Closed-Cup Flash Point Apparatus is designed and manufactured as per standards GB/T 261-2008
Determination of Flash Point - Pensky-Martens Closed Cup Method and
ASTM D93 Standard Test Method for Flash Point by Pensky-Martens
Closed Cup Tester. It is used to make determination of the closed
cup flash point of the petroleum products.
I. Main technical features
1. Manual Pensky-Martens Closed-Cup Flash Point Apparatus is designed as per GB/T261 and ASTM D93.
2. The heating power is continuously adjustable. The power is shown
visually by voltmeter.The temperature control mode is advanced and
reasonable.
3. The structure is designed small and exquisite.The
stainless-steel table board is beautiful and easy to clean.
4. Operation is easy.Test results are precise.It has been chosen as
the instrument to test the closed cup flash point by many metering
and detecting institutes.
II. Main technical specifications
1. Power supply:AC (220±10%)V, 50Hz.
2. Heating device::The heating power is continuously adjustable
from 0 W to 600W.
3. Heating rate::(1~12)℃/min;Controllable and adjustable.
4. Stirring rate: Procedure A: (90~120)RPM, Procedure B:
(250±10)RPM
5. Oil cup:(1) Inner diameter: 50.7mm~50.8mm.
(2) Depth: 55.7mm~56.0mm
(3) The scribed line depth of capacity of testing oil:
33.9mm~34.3mm.
(4) Capacity of testing oil:about 70ml
6. Igniting device:(1) Igniting source: gas(or other civilian
fuels,the same below)
(2) Electric ignition. Gas flame diameter:3.2mm~4.8mm
7. Thermometers:Mercury-in-glass thermometer.Specifications are as
below:
(1)Scale -5℃~110℃, division 0.5℃.
(2)Scale 20℃~150℃, division 1℃.
(3)Scale 90℃~370℃,division 2℃.
8. Ambient temperature:≤ 35℃
9. Relative humidity:≤ 85%
10. Maximum power consumption:650W
III.Accessories
No. | Item | Unit | Qty | Remarks |
1 | Thermometer -5 ~ 110 ℃, each scale division is 0.5 ℃ | Piece | 1 | GB / T261-2008 Standard requirement |
2 | Thermometer 20 ~ 150 ℃, each scale division is 1.0 ℃ | Piece | 1 |
3 | Thermometer 90 ~ 370 ℃, each scale division is 2 ℃ | Piece | 1 |
4 | Driving flexible shaft | Piece | 1 | |
5 | Latex tube Ф3 250mm | Piece | 3 | One for alternate |
6 | 5A fuse (Ф5 × 20) | Piece | 2 | |
7 | Thermometer adapter | Set | 1 | |
8 | Thermometer liner (Cut suitable length of liner to wrap the lower end of the
thermometer, so that the thermometer is just inserted into the
thermometer adaptor) | Piece | 5 | |