Add to Cart
Hexavalent chromium Single Beam Spectrophotometer Indigo UV Photometer
Competitive Advantage:
● UV-1800S series dual-beam optical system, successfully achieved
the perfect combination of high precision and high reliability of
the measurement can meet the requirements of various applications,
can be used in biological research, biotechnology industry,
pharmaceutical analysis, pharmaceutical, teaching and research,
environmental protection , food hygiene, clinical testing, health
and epidemic prevention and other areas
● wide wavelength range, which can meet various areas of the
wavelength range of requirements
● UV-1800S has strict requirements 0.5 / 1.0 / 2.0 / 4.0nm four
spectral bandwidth to meet the pharmacopoeia
● Automatic design, to achieve the most simple means of measurement
● LSI design greatly improves the scalability and reliability of
the system
● Improved optimization of optical design, import source and
receiver creating a system of high performance and high reliability
● Rich measurement methods, with a wavelength scan, time scan,
multi-wavelength determination, multi-order derivative
determination (optional), dual-wavelength, three wavelength
(optional) DNA protein measurement (option) and other measuring
methods, can meet different measurements requirements, and can be
displayed directly on a 6-inch screen
● based on user requirements optional hole rack, manual four with
rack, manual Eighth rack, automatic Eighth frame, glass frame, test
tube rack, 1cm colorimetric frame, 5cm colorimetric frame, 10cm
colorimetric racks
● Measurement data can be output through the printer with USB
interface
● can be powered down to save the measurement parameters and data,
user-friendly
● Realize spectral scanning by PC software control, more precise
and flexible measurement requirements
Product Introduction:
UV-visible spectrophotometer UV-1800S, 4nm, 2nm, 1nm, 0.5nm spectral bandwidth four custom installation according to user requirements, can meet the stringent requirements of pharmacopoeia
Specifications:
Wavelength range: | 190~1100nm |
Spectral Bandwidth: | 0.5/1.0/2.0/4.0nm |
Wavelength accuracy: | ±0.3nm |
Wavelength Reproducibility: | ≤0.1nm |
Transmittance accuracy: | ±0.3%τ (0-100%τ) ±0.002A(0~0.5A) ±0.003A(0.5A~1A) |
Transmittance Repeatability: | ±0.15%τ (0-100%τ) ±0.001A(0~0.5A) ±0.0015A(0.5A~1A) |
Stray light: | ≤0.03%τ(220nm NaI,340nm NaNo2) |
Stability: | 0.0005/h(500nm Preheated) |
Metering: | Automatic |
Wavelength tuning: | Transmittance, absorbance, concentration, energy |
Photometric range: | -4~4A |
Display: | 6 inches high brightness LCD display |
Detector: | Imports of silicon photodiode |
Source: | Imported deuterium lamp, tungsten lamp imports |
Power supply: | AC 220V/50Hz,110V/60Hz |
Power: | 120W |
Instrument Size: | 560×450×230mm |
Net weight: | 28Kg |
About Image:
Applications:
Drug analysis: | National Formulary" Available UV-visible spectrophotometry drugs for drug testing, drug analysis, pharmaceutical and other industries |
Life Sciences: | DNA, protein concentration |
Animal husbandry and fishery: | Pesticide residues, crop composition testing, veterinary testing, feed testing, fertilizer testing, soil composition testing, testing and other aquaculture |
Metal elements: | Ag Cu Pb Zn Co Ni K Na Ca Mg Li Mn Cr As, including precious metals |
Geological Survey: | Determination of metal elements in minerals and inorganic salts |
Food safety: | Content of additives, preservatives, perfumes, fats, enzymes, sugars, spices, minerals, vitamins, etc. |
Environmental monitoring: | Monitoring water, air, rainfall and soil, etc., in which the determination of various pollutants |