Add to Cart
Features:
• 6 Standard parameters: ECG, TEMP, NIBP, SpO2, RESP, PR
• 8 inch high resolution color screen
• Real time S-T segment analysis, pace-maker detection
• Indication of incorrect manipulation and malfunction analysis
• Patient info input management function
• Large volume of tabular and praphic trends information storage and easy to recall
• Capture 2-hour dynamic trend
• Various interfaces:standard screen,trend screen,oxyCRG screen,big front screen
• Up to 400 groups NIBP list,60 alarm event records recall, 7-day trend chart in storage.
• Up to 2 hours working capacity of built-in rechargeable
lithium battery
• Unique accessory management cabinet
• Networking capacity
• ETCO2, cart, wall shelf are optional
Parameters:
ECG | Lead Mode 3-lead and 5-lead are optional Lead Selection I, II, III, avR, avL, avF, V Wave 5-lead: 2 channels 3-lead: 1channel Gain ×2.5mm/mV, ×5.0mm/mV, ×10mm/mV, ×20mm/mV HR Measuring and Alarm Range Range 15 ~ 300 bpm Accuracy ±1% or ±1bpm, which is greater Alarm Accuracy ±2bpm Resolution 1 bpm |
ST Segment Monitoring | Measuring and Alarm Range -0.6 mV~ + 0.8 mV |
Alarm | Available |
Review | Available |
Scan Speed for ECG Waveform is adjustable | 12.5mm/s accuracy ±10% 25mm/s accuracy±10% 50mm/s accuracy±10% |
Respiration | Method R-F(RA-LL) Impedance Differential Input Impedance >2.5 MΩ Measuring Impedance Range 0.3~5.0Ω Baseline Impedance Range 100Ω– 2500Ω Bandwidth 0.3 ~ 2.5 Hz |
Resp. Rate | Measuring and Alarm Range 0~120rpm Resolution 1 rpm Measuring Accuracy ±2 rpm Alarm Accuracy ±3rpm Apnea Alarm 10 ~ 40 S |
NIBP | Method Oscillometry Mode Manual, Auto, continuous Measuring Interval in AUTO Mode 1 /2 /3 /4 /5/ 10/ 15/ 30 /60 / 90/120/240/480/960 Min Measuring Period in Continuous Mode 5 Min Measuring and Alarm Range 10~270mmHg Alarm Type SYS, DIA, MEAN |
Resolution | Pressure 1mmHg Cuff Pressure ±3 mmHg Accuracy ±10% or ±8mmHg, which is greater Over-pressure Protection: Adult Mode 315±10 mmHg Pediatric Mode 265±10 mmHg Neonatal Mode 155±10 mmHg |
SpO2 | Measuring Range 0 ~ 100 % Alarm Range 0 ~ 100 % Resolution 1 % Accuracy 70% ~ 100% ±2% 0% ~ 69% unspecified |
Pulse Rate(PR) | Measuring and Alarm Range 0~250bpm Resolution 1bpm Measuring Accuracy ±2bpm or ±2%, which is greater Alarm Accuracy ±2bpm |
TEMP | Channel dual-channel Measuring and Alarm Range 0 ~ 50°C Resolution 0.1°C Accuracy ±0.1°C Actualization Interval about 1 Sec. Average Time Constant < 10 Sec. Alarm responding Time ≤2min |
Packaging Details: | Packing: 1 unit/carton 1* Patient monitor reusable Spo2 sensor (1 pc); ECG cable and leadwire (1 pc); NIBP cuff and tube (1 pc); Built-in lithium battery (1 pc) 1*English user manual (1pc) |
Delivery Detail: | Shipped in 7 days after payment |