ECG simulator 10 leads ECG machine simulator with 2 type waveform patient monitor simulator build-in battery
Feature of medical simulator:
1.Rechargeable Lithium battery |
2.10 pcs multi-functional ECG connector, can be connect with different type ECG machine cable or patient monitor cable |
3.Battery power indicate light |
4.Low battery power automatic turn off function, protect the Lithium battery |
Specification of simulator:
The simulator can generate 2 types waveform
1.Normal ECG waveform
2.Plus-minus triangular waveform
Dimension: 15x10x3cm
N.W.: 0.35kg
Note:
1.The simulator will automatically generate Normal ECG waveform
when connect to power
2.The LED display will be off automatically if without operation
after 5 seconds, press any key will be returned to display
Key Description:
There are 4 keys, followed by selection keys, increase key,
reducing key, the Enter key, in addition to a set of key
combinations
Select Key:
This key is used to select the parameter you want to change; there
are 4 LED tubes to display the four codes represent what is
displayed,
The 1st LED representative waveform code, LED 2-4 representatives
the parameter (2 is the value of one hundred, 3 representatives of
ten, 4 representatives of bits)
Operating method described separately for each waveform:
1. Normal ECG waveform:
Heart rate setting range: 10-200bpm; (initial value: 60 bpm)
Signal amplitude fixed.
2. Plus-minus triangular waveform:
Frequency range: 10-300bpm; (initial value: 75 bpm)
Amplitude range: 0.1-4Mv (10: 0.1mV, 400: 4mV);
2 modes, Mode 1: Positive waveform; mode II: negative waveform; Use
combinations key to choose.
Note: The signal amplitude settings:
amplitude range: 0.1mV-4mV (10: 0.1mV, 400: 4mV); (initial value:
100)
Please note that this amplitude setting will affect all waveforms.
Please note that the signal amplitude settings do not affect the
amplitude of the ECG waveform.