Safety Testing Medical Equipment and Technology for medical device manufacturers and governmental quality department

SVJ-Y2 Medical Residual Voltage Tester

Specification:

Test Voltage: adjustable from 0V to 250V (AC); Displayed on the touch screen; error: within ±2% of reading

Measuring Range of residual voltage: 0~60V; Displayed on the touch screen; error: within ±1V

Displayable range of residual voltage: 0~242V

Maximum residual voltage is collected instantaneously with immediate touch - screen display.

Alarm threshold: if residual voltage exceeds 60V, the tester will sound an alarm automatically.

Method of testing: power on time, power off time and times of testing can be set on the touch screen, PLC realizes automatic cycle of the test

Power: 2KW