Waveform B Scan Color Screen Wall Ultrasonic Thickness Gauge Single And Dual Element

Brand Name:HUATEC
Certification:CE ISO
Model Number:TG-5600DL
Minimum Order Quantity:1
Place of Origin:China
Supply Ability:100 per months
Contact Now

Add to Cart

Verified Supplier
Location: Beijing Beijing
Address: 7th Floor, Chengyuan Building B, the Mid. Road of Jiancaicheng Haidian Dist.BeiJing,CHINA 100096
Supplier`s last login times: within 48 hours
Product Details Company Profile
Product Details

Waveform B Scan Single And Dual Element Color Screen Wall Ultrasonic Thickness Gauge


HUATEC TG-5600DL is a portable ultrasonic thickness gauge. Based on the operating principles of ultrasonic measurement, TG-5600DL is capable of measuring the thickness of various materials with accuracy as high as 0.001 millimeters or 0.0001 inches, and widely used in a variety of thickness measurement applications.

Users can measure the thickness from only one side of the material being measured, without damaging the workpiece.


TG-5600DL can display thickness readings and A-scans at the same time, allowing users to check waveforms. The echo display helps users control the measurement, so that the thickness measurement range is large, the measurement values are correct, and material-related errors in measurement can also be avoided. The B-scan function is used to observe the bottom profile of the workpiece being measured. The probe moves along the surface of the workpiece to display the cross-section of the workpiece.


This gauge can be used in manufacturing, metal processing industry, aerospace, railway transportation, chemical industry, commodity inspection industry and other testing fields.In addition to accurately measuring a variety of plates and processed parts, monitoring a variety of pipes and pressure vessels that in production equipments to further monitor their thinning levels after being corroded during use, which is an indispensable tool in the non-destructive testing industry.

The storage function is optional, according to the different needs of customers, and built-in large-capacity data logger can store up to 100,000 thickness measurement values and 400 waveforms. Other more functions can be optional, such as multi-layer thickness measurement, ultra-high resolution, large-range thickness measurement,digital noise reduction and window mode, inner wall oxide layer, real-time communication, multi-channel and other options.


Main Measurement Functions

• Most material measurements can be made using single element transducers.

• Measuring thickness range: 0.01 mm ~ 38 m, depending on the material being measured, probe, surface conditions, temperature and selected configuration.

• Determine the sound velocity of an unknown sound velocity material

• High resolution: Regular display resolution is 0.001mm, optional ultra-high resolution can reach 0.0001mm.

• Measurement units: millimeters, inches, microns and mils.

• High temperature measurement: temperatures up to 500°C.

• System default configuration menu: For user convenience, it has been set according to the default conditions for using the probe of the instrument.

• Pause function: allows post-measurement processing operations. Conveniently adjust all parameters to obtain stable measurement data.

• Full-screen waveform display function: increases the echo display area by 20%, giving the operator a large window experience.

• Echo intensity indication: The stronger the echo, the more indicator bars are displayed, especially the digit display, which is more useful.

• Time-based B-scan mode: It is used to observe the bottom profile of the workpiece to be measured.

• Difference/Reduction Rate Mode:The thickness difference relative to the set point is displayed and expressed as a percentage of the reduction rate.

• TDG time-varying gain: The initial pulse or interface echo can be effectively reduced by reducing the initial gain, and the bottom echo of the tested sample can be increased by increasing the maximum gain, which is easy to measure.

• Measurement Mode:Pulse-Echo(P-E),Interface-Echo(I-E),Echo-Echo(E-E),Multiple echo(I-ME),Double echo verification(E-EV)


Main Parameters

Instrument Specifications
Working principleP-E (Pulse-Echo), I-E (Interface-Echo), E-E (Echo-Echo), I-ME(Multiple Echoes), E-EV(Echo verification)
Probe TypeDelay line probe, Contact probe, Membrane probe, Pencil type probe, Immersion probe
CalibrationProbe zero, Calibrate velocity, 1-Point calibration, 2-Point calibration, Set Velocity
Thickness Range

0.01mm~38m,Depending on the tested material, probe surface conditions, temperature, and selected configuration.

0.15mm~30mm(Standard D15A probe,steel)

Material velocity range310 - 18699m/s,and can measure the velocity with known thicknes
ResolutionOptional 0.1/ 0.01/0.001mm
Probe Frequency0.5MHz~20 MHz
Measuring Unitmm, inch,μm,mil
Update Rate4HZ, 8HZ, 16HZ per second optional
Power OffAutomatically power off when idle for 5,10 or 20 minutes, or manually
Display Type2.4 in (320 × 240 dot matrix)TFT color display LCD screen
Rectification ModeRF, RF-, FULL, HALF+, HALF
View ModeA-SCAN, B-SCAN, DIGITS, CAPTURE, DIFFERENCE, ALARM
StorageCan store 100,000 thickness values and 400 parameter setting files(Optional)
GainLow, Medium or High for varying test conditions
Display ModeNormal, Minimum / Maximum capture, DIFF/RR%
V-Path CorrectionAutomatic
Update RateSelectable 4Hz, 8Hz, 16Hz
Material Velocity Range500 to 9999m/s (0.0197 to 0.3937in/us)
LanguagesSelectable Chinese, English, Japanese
Alarm SettingsMinimum and Maximum alarms. Range of 0.25 mm to 508 mm (0.010" to 20.00"). Dynamic waveform color change on alarm
Power Requirements2 AA size batteries
Operating TimeApproximately 40 hours
Instrument Shut-offSelectable ALWAYS ON or AUTO OFF after 5, 10, 20 minutes of inactivity
Operating Temperature-10°C to +50°C (+10°F to +120°F)
Size153mm X 76mm X 37mm(H X W X D)
Weight280g including batteries

Standard Configuration

ItemsQuantity

TG-5600DL

Ultrasonic Thickness Gauge

1
Probe1
Cable1
Rubber Sheath1
Alkaline Batteries-
Couplant Bottle1
Case1
Operation Manual1
Calibration Certificate1

Optional accessories

ProbesCables
Rubber SheathStep Blocks
Delay LinesCoupling Agent
Protective filmSoftware Options

Appendix A: Material Velocity

The table lists the propagation velocity of ultrasond in various common materials. It is only for reference. The actual velocity of these materials may differ from the velocity in the table for various reasons, for example, material composition, preferred orientation of grains, porosity, and temperature etc. Therefore, to achieve the best measurement accuracy, a sample made of a specific material should first be tested to confirm the velocity of specific material.

Table:Material Velocity List

MaterialVelocity(m/s)
Steel 20#5891
Aluminum6320
Chromium6200
Copper4660
Gold3251
Iron5930
Lead2160
Magnesium5806
Silver3607
Titanium6098
Zinc4193
Tungsten5181
Tin3320
Brass4380
Grey Cast Iron4600
Glass5766
Nylon2610
Stainless Steel 3045334
PVC2392
Silicon9620
Quartz Glass5655

Note:Above table is for reference only. For actual velocity calibration, refer to 6.2 velocity calibration