1. Test object: home appliance production line, laboratory
2. Test items: Hipot, grounding, leakage current, insulation resistance, power
3. Technical specification
AC Hipot | |
Output voltage setting range/ precision | AC 200~5000V ±(3%×setting value+5words) |
Breakdown current testing range/ precision | 0.10~25.00mA ±(3%×setting value+5words) |
Breakdown current presetting alarm range | Max:0.10~25.00mA ;Min:0.00~5.00mA |
Test time range/ resolution | 1~300s 1s/step |
Grounding resistance | |
Grounding current output range/ precision | AC 5.0~30.0A ±(3%×setting value+5words) |
Grounding resistance alarm setting range | Max:5~10A:10.0~600mΩ;11~25A:10.0~300mΩ;26~30A:10.0~200mΩ; Min:0.0~100mΩ |
Grounding resistance test range/ precision | 10.0~600mΩ ±(3%×display value+5words) |
Test time/ resolution | 1~60s 1s/step |
Leakage current | |
Output voltage | 1.06 times rated voltage (external provided) |
Leakage type setting | Dynamic leakage, static leakage, optional equipped with 3 types of body network, which can be customized and switched freely. |
Output voltage setting range/precision | AC 50~280V ±(0.5%×display value+2 words) |
Leakage current setting range/precision | 0.05~20.00mA ±(3%×display value+10uA) |
Leakage current setting alarm range | Max.: 0.05~20.00mA; Min.: 0.01~5.00mA |
Test time range/resolution | 1~300s 1s/step |
Insulation resistance | |
Output voltage setting range/ precision | DC 500/1000V ±(3%×reading value+5 words) |
Insulation resistance testing range | 1~2000MΩ ≤100MΩ:±(5%×display value +3 words) >100 MΩ:±(8%×display value +8 words) |
Insulation resistance alarm setting range | Max:0~2000MΩ;Min:1~2000MΩ |
Test time range/ resolution | 1~300s 1s/step |
Power | |
Output voltage | 1.0 times rated voltage |
Output voltage measurement range/precision | AC 50~280V ±(0.5%×display value+2 words) |
Output current measurement range/precision | AC 0.20~30.00A ±(0.5%×display value+2 words) |
Active power measurement range/precision | 30~6000W ±(0.5%×display value+5 words) |
Power setting alarm range | Max.: 3~6000W Min.: 0~6000W |
Test time range/resolution | 1~300s 1s/step |