Product Details

IEC60695-11-3 UL94 Horizontal-Vertical Flame Chamber
Product information
Horizontal and Vertical Flammability Testing Machine
UL 94 flame chamber. Horizontal flame chamber, vertical flame chamber
For the flammability testing of plastic materials used in consumer electric devices and appliances. Stainless steel construction with black interior and large sliding window for easy viewing. Includes a fully adjustable specimen support and precision gas flow controls.


Technical parameters

Kinds500W(5V class )50W(HB/Vclass)
Test Flame500W(burning test for V-0,V-1,V-2 classification)50W(burning test for V-0,V-1,V-2 classification)
Burner TubeInternal diameter: Φ 9.5mm±0.3mmInternal diameter: Φ 9.5mm±0.3mm,
Burning Angle0°,20°,45°0°,20°,45°
ThermocoupleType K (Ni/Cr - Ni/Al)Type K (Ni/Cr - Ni/Al)
Thermocouple Size0.5 mm0.5 mm
Standard Copper Block

9mm±0.01mm,
weight:10g±005g before drilling.

5.5mm±0.01mm,
weight:76g±0.01g before drilling.

Time for Temperature to IncreaseFrom 100°C±5°C to 700°C±3°C; 54s±2s (preset)

From 100°C±5°C to 700°C±3°C;
44s±2s (preset)

Gas Flow965±32ml/min105±10ml/min
Gas Supplymethane,min.98% puritymethane,min.98% purity
Water-column125m±5mm‹10mm
Test Time54±2s44±2s
Flame Height Indicator40±1mm20±1mm
Volume Of Chamber>0.5m³, black interior>0.5m³, black interior
Conform to standardUL94,IEC60695-11-3,IEC60695-11-20UL94,IEC60695-11-4,IEC60695-11-10


Testing Standards
UL 94-2006 IEC 60695-11-3 IEC 60695-11-4

UL94 Horizontal & Vertical Flammability Tester.pdf...


Specimen holder


Controll panel