Add to Cart
4.6 Kg 150 * 540 * 380 mm Portable X Ray Imaging Scope For Hospital Orthopedics
Description:
Low-Intensity Portable X-Ray Imaging Scope (the following in short X-Ray Imaging Scope), a new type of X-Ray medical instruments in the world, consists of main engine and power supply. From the appearance, the main engine adopts the structure of C-Type-Arm, which amplifies the radius of observation. At the same time, the materials select ABS engineering plastics, heightening the insulation capacity.
Because of adopting MCP-X-Ray imagine strengthening device, having compact texture, high gain, high resolution, the scope can produce clear real-time images without darkroom. The advance technology of widening pulse is used to control electrical appliance of X-Ray tube, which has many measures to be protected, so it has high working stability and no noises.
In addition, the scope takes strict precautions, the leak of X-Ray reducing, and then the operators do not need any protections. This scope applies to orthopedics and pediatrics in the hospitals and to diagnose IUD for childbearing-age women in time. Especially, it applies to various field rescues in such as industry materials and apparatus quality inspection, athletes, armies, field working and safety checking, and so on.
The scope belongs toⅠClass B Model Equipment.
The technical performance parameters:
Effective field of view 75*100mm
Thickness of measured object ≤300mm
Tube voltage 45~90kv continuously adjustable
Stabilized voltage precision ≤±10%
Tube current 0.25~0.5mA continuously adjustable
Focus position and reference axis deviation ±1㎜
Stabilized current precision ≤±20%
Brightness ≥25cd/m2
Contrast 1000:1
Image resolution ≤3lp/mm
Gray scale ≥7 level
LEAKGE RADIATION IN THE LOADING STATE
In the LOADING STATE,the AIR KERMA due to LEAKAGE RADIATION from X-RAY TUBE ASSEMBLIES and X-RAY SOURCE ASSEMBLIES, at 1m from the FOCAL SPOT,averaged over any area of 100cm2 of which no principal linear dimension exceeds 20cm,when operated at the NOMINAL X-RAY TUBE VOLTAGE conditions of LOADING conditions,shall not exceed 1.0 mGy/h .Particular standards may specify different limits of LEAKAGE RADIATION for specific X-RAY EQUIPMENT.
Environmental conditions
a)Ambient temperature range +10℃~+40℃
b)Relative humidity Range 30%~75%
c) Atmospheric pressure range 700hPa~1060hPa
The x-ray tube parameters
Nominal focal spot value: 0.8 mm Maximum operating voltage: 90 kV
Maximum Power Current: 1.0 mA Anode heat content: 4 kJ
Storage conditions and methods:
1. Store ventilation and dryness to prevent dampness.
2. Transport and storage of environmental constraints:
a) Ambient temperature range:+10℃~+40℃;
b) Relative humidity Range: 30%~75%;
c) Atmospheric pressure range:700hPa~1060hPa;
Strong Packing: